The X-STREAM Enhanced XEGP general purpose continuous gasanalyzer measures up to five components simultaneously in various combinations. Its versatile design allows physical benches to be installed in their own compartment, separate from the electronics and comes in a 19 inch rack mount or table top enclosure.
Power Supply | 100–240 V , 50/60 Hz |
Analog Input | 2 analog inputs, 0-1(10) V (Rin = 100 kΩ) or 4(0)-20 mA (Rin = 50 Ω) |
Analog Output | 1-5, individually optically isolated 4(0)-20 mA (RB ≤ 500 Ω) |
Relay Outputs | 4 status relays acc. NAMUR NE 107 or e.g. concentration thresholds, valve status notification dry contacts: 1 A, 30 V |
Compliances | EN 61010-1, EN 61326, NAMUR, CSA-C/US, C-Tick |
Gas connections | PVDF: 6/4 mm; Stainless steel: 6/4 mm or 1/4 in.; for more options c.f. |
Rated voltage | 100–240 V , 50/60 Hz |
Rated input current | 3–1.5 A |
Power input | IEC mains appliance |
Signal connections | Submin connectors or screw terminals; RJ45, USB |
Enclosure protection | IP 20 acc. EN 60529 for indoor installation, protected against direct sunlight |
Humidity | < 90 % r.h. @ 20 °C (68 °F) < 70 % r.h. @ 40 °C (104 °F) |
Communication Protocol | Ethernet with Modbus TCP; RS 485 / 232C with Modbus RTU; 2 USB ports |
Rosemount™ X-STREAM Enhanced XEGP Continuous Gas Analyzer